O2 Lite PPO2 Hyperbaric Monitor

Description

The model O2-Lite hyperbaric portable oxygen monitor measures oxygen partial pressure with a long life electro-chemical sensor.

Oxygen partial pressure is displayed with 0.001 atm/bar O2 resolution on a back-lit LCD display. The measurement is compensated for ambient temperature variations.

The monitor has an adjustable high and low audio visual alarms with mute.

Possible applications include:

  • Diving Bell ppO2 monitoring
  • Hyperbaric chamber ppO2 monitoring

Features

  • Designed for hyperbaric use
  • Tested to 350 MSW in heliox
  • Compliant with IMCA D023 & D024
  • Lloyds design appraisal for use to 350 MSW
  • Single point calibration in ambient gas or calibration gas
  • Can be used as a portable or panel mounted using an optional bezel
  • Powered from 24VDC or its internal rechargeable battery that will provide about a month autonomy between charges
  • Battery status indication and low battery warning
  • Legend can be changed from atm to bar O2 by the user

Specifications

Range: 0 to 2.00 atm/bar oxygen
Resolution: 0.001 atm/bar oxygen
Accuracy: ± 0.005 atm/bar O2
Sensor Life: ~5000 atm/bar O2-hours, ~2.5years at 0.2 atm/bar O2
Power Supply: 18 to 28 VDC and 6 Volt NIMH rechargeable battery
Environmental: Sealed to IP65
Dimensions: 120mm x 80mm x 60mm
Weight: 350 grams
Model Number: A08-35-15

Request Quotation
Request a Quotation

Thank you for your interest in receiving a Quotation for our instrumentation.


Below you will find three short steps to follow in order for us to produce a quotation. To enable us to deal with your request efficiently, please ensure that you supply your full Company details together with your exact requirements.

Please note: all quotations are manually generated based on the information you will supply. Details of your intended supply voltage (if applicable) will ensure we supply the correct product code and lead time.
step 1 step 2 step 3
Please register on our Helpdesk, supplying your full company details Please check the email account that you used when registering, we have sent you an email with a link to activate your account Once your account is activated - Login to our helpdesk and select Request and Quotation
Please Click Here To Register Now! If you do not receive our activation email or experience any problems registering, please Contact Us and we will manually create your account. Login To Our Helpdesk Now!

dive communication solutions
Request a quotation on our instrumentation data logging systems or communications equipment data logging spares and accessories
S.A.T.Systems Ltd   1 Kelly Industrial Estate, Kirk Michael, Isle of Man,  IM6 1SS